Skip to main content

Research Repository

Advanced Search

Periprosthetic femoral fractures around the original cemented polished triple-tapered C-stem femoral implant: a consecutive series of 500 primary total hip arthroplasties with an average follow-up of 15 years

Baryeh, K; Wang, C; Sochart, D

Authors

K Baryeh

C Wang

D Sochart



Abstract

IntroductionThe true incidence of periprosthetic femoral fracture (PFF) around cemented polished taper-slip implants remains largely unknown. Registries usually only capture PFFs that result in revision, missing those managed non-operatively or treated by open reduction and internal fixation (ORIF). This study reports the long-term rate of PFF with the original triple-tapered C-stem femoral implant.Materials and methodsA prospective review of a consecutive series of 500 primary total hip arthroplasties (THAs) performed at a single centre between March 2000 and December 2005, with average follow-up of 15 years (12-19 years).ResultsThere were 500 consecutive THAs in 455 patients. Seven PFFs (1.4%) occurred in seven patients at an average of 7.9 years (range 2-11.5) from the primary arthroplasty. Five PFFs were managed by ORIF, one Vancouver B3 fracture was revised for a loose implant and one patient was treated non-operatively. Average age at primary operation was 74 years (67-87) and BMI averaged 27.3 (22-31). There was no typical fracture pattern and no statistically significant associations with patient demographics (age, gender, BMI, diagnosis) or prosthetic details (size, offset, alignment, cement mantle, subsidence). Survivorship to the occurrence of PFF was 99% (97.3-99.6%) at 10 years and 97.8% (95.5-99.0%) at 15.ConclusionA PFF rate of 1.4% at an average follow-up of 15 years represents the true incidence of PFF with the use of the original triple-tapered C-Stem femoral implant, similar to that of published Exeter series (1.85%) but lower than the CPT (3.3%).

Citation

Baryeh, K., Wang, C., & Sochart, D. (2022). Periprosthetic femoral fractures around the original cemented polished triple-tapered C-stem femoral implant: a consecutive series of 500 primary total hip arthroplasties with an average follow-up of 15 years. Archives of Orthopaedic and Trauma Surgery, https://doi.org/10.1007/s00402-022-04712-x

Journal Article Type Article
Acceptance Date Nov 22, 2022
Online Publication Date Nov 29, 2022
Publication Date Nov 29, 2022
Deposit Date Jan 12, 2023
Journal Archives of Orthopaedic and Trauma Surgery
Print ISSN 0936-8051
Electronic ISSN 1434-3916
Publisher Springer Verlag
DOI https://doi.org/10.1007/s00402-022-04712-x
Publisher URL https://doi.org/10.1007/s00402-022-04712-x


Downloadable Citations