Optimising image quality and radiation dose for neonatal incubator imaging

Tugwell-Allsup, J; Morris, R; Gibbs, R; England, A

Optimising image quality and radiation dose for neonatal incubator imaging Thumbnail


Authors

J Tugwell-Allsup

R Morris

R Gibbs

A England



Abstract

Introduction
Neonates often require imaging within incubators however limited evidence exists as to the optimal method and acquisition parameters to achieve these examinations. This study aims to standardise and optimise neonatal chest radiography within incubators.

Methods
A neonatal anthropomorphic phantom was imaged on two different incubators under controlled conditions using a DR system. Exposure factors, SID and placement of image receptor (direct v tray) were explored whilst keeping all other parameters consistent. Image quality was evaluated using absolute visual grading analysis (VGA) with contrast-to-noise ratio (CNR) also calculated for comparison. Effective dose was established using Monte Carlo simulation using entrance surface dose within its calculations.

Results
VGA and CNR reduced significantly (p < 0.05) whilst effective dose increased significantly (p < 0.05) for images acquired using the incubator tray. The optimal combinations of parameters for incubator imaging were: image receptor directly behind neonate, 0.5 mAs, 60 kV at 100 cm SID, however, if tray needs to be used then these need to be adapted to: 1 mAs at maximum achievable SID. Effective dose was highest for images acquired using both incubator tray and 100 cm SID owing to a decrease in focus to skin distance. There is significant increase (p < 0.01) in VGA between using 0.5 mAs and 1 mAs but an apparent lack of increase between 1 and 1.5 mAs.

Conclusion
Using the incubator tray has an adverse effect on both image quality and radiation dose for incubator imaging. Direct exposure is optimal for this type of examination but if tray needs to be used, both mAs and SID need to be increased slightly to compensate.

Implications for practice
This study can help inform practice in order to both standardise and optimise chest imaging for neonates in incubators.

Citation

Tugwell-Allsup, J., Morris, R., Gibbs, R., & England, A. (2020). Optimising image quality and radiation dose for neonatal incubator imaging. Radiography, 26(4), e258-e263. https://doi.org/10.1016/j.radi.2020.03.011

Journal Article Type Article
Acceptance Date Mar 23, 2020
Online Publication Date Apr 9, 2020
Publication Date Nov 1, 2020
Deposit Date Apr 29, 2020
Publicly Available Date Apr 9, 2021
Journal Radiography
Print ISSN 1078-8174
Electronic ISSN 1532-2831
Publisher Elsevier
Volume 26
Issue 4
Pages e258-e263
DOI https://doi.org/10.1016/j.radi.2020.03.011
Publisher URL https://doi.org/10.1016/j.radi.2020.03.011
Related Public URLs https://www.journals.elsevier.com/radiography

Files




You might also like



Downloadable Citations