Skip to main content

Research Repository

Advanced Search

Estimating the error of CT-based measurements of aortic lumen volume used in endovascular planning

England, A; Fisher, R K; McWilliams, R G; Torella, F

Authors

A England

R K Fisher

R G McWilliams

F Torella



Abstract

Introduction
Planning of endovascular sealing of abdominal aortic aneurysms requires measurement of the aortic lumen volume. The aim of this study was to investigate mathematically the effect of intra- and inter-observer variability error, as well as cardiac cycle-related variability, on these measurements.
Methods
Mean (±2SD) intra- and inter-observer error in lumen measurements and mean (+2SD) cardiac cycle-related variability were obtained from published literature and added to the measurement of the flow lumen volume of a 57 mm abdominal aortic aneurysm to calculate average and extreme error possibilities.
Results
The aneurysm volume was measured at 165 ml. The calculated possible mean measurement error due to cardiac cycle variation, intra- and inter-observer variability was +11.0%, resulting in a potential measurement of 183.1 ml. The calculated extreme errors were +24.3% (if 2SD of all errors were added to the mean) and +3.5% (if 2SD of all errors, except cardiac cycle, were subtracted from the mean), resulting in potential measurements of 170.8 ml and 205.1 ml, respectively. When considering the errors combined, the proportion of patients who may have volume measurement errors of up to ±2.5 ml, ±2.6 to ±5.0 ml and ±5.1 to ±7.5 ml were 18%, 17% and 15%, respectively.
Conclusion
Measurement of CT-based aortic lumen volumes in abdominal aortic aneurysms is imprecise. This has practical implications for the planning and the performance of complex endovascular therapies.

Citation

England, A., Fisher, R. K., McWilliams, R. G., & Torella, F. (2017). Estimating the error of CT-based measurements of aortic lumen volume used in endovascular planning. Radiography, 23(4), 287-291. https://doi.org/10.1016/j.radi.2017.08.001

Journal Article Type Article
Acceptance Date Aug 4, 2017
Online Publication Date Sep 1, 2017
Publication Date Sep 1, 2017
Deposit Date Nov 30, 2017
Journal Radiography
Print ISSN 1078-8174
Publisher Elsevier
Volume 23
Issue 4
Pages 287-291
DOI https://doi.org/10.1016/j.radi.2017.08.001
Keywords Aorta, CT, Endovascular, Errors, Planning, Variability
Publisher URL http://dx.doi.org/10.1016/j.radi.2017.08.001
Related Public URLs http://www.radiographyonline.com/