AP versus PA positioning in lumbar spine computed radiography : Image quality and individual organ doses
Davey, E; England, A
Authors
A England
Abstract
Purpose: Radiological imaging examinations must be optimised in order to ensure that the radiation dose is kept as low as reasonably possible (ALARP). The aim of this study was to compare anteroposterior (AP) and posteroanterior (PA) projections of the lumbar spine, at various kVp increments, in order to establish optimum parameters.
Methods: An anthropomorphic phantom was imaged in both the AP/PA projections and at various kVp increments. Acquisitions were undertaken using a Wolverson Acroma X-ray unit and processed using an Agfa computed radiography (CR) unit. The entrance surface dose was recorded and converted to effective and organ doses using PCXMC 2.0 software. Five observers were then asked to evaluate the images, using a two-alternative force choice (2AFC) approach and a scale based on EC guidelines.
Results: The PA projection lowered the mean effective dose by 19.8% and also the mean absorbed dose to the stomach (70.4%), colon (61.1%), remainder tissues (33.2%), ovaries (7.3%) and testes (15.9%). However, this was at the expense of slightly inferior image quality, not statistically significant. For AP projections, a higher kVp is a further option for dose reduction.
Conclusion: Dose optimisation requires the production of an image that is acceptable for the purpose intended. Based on ALARP, and when taking into consideration the dose reductions in this study, it may now be time to routinely use PA projections when imaging the lumbar spine. The use of a higher kVp should also be considered as an option but would be more useful for AP projections.
Citation
Davey, E., & England, A. (2015). AP versus PA positioning in lumbar spine computed radiography : Image quality and individual organ doses. Radiography, 21(2), 188 - 196. https://doi.org/10.1016/j.radi.2014.11.003
Journal Article Type | Article |
---|---|
Acceptance Date | Nov 18, 2014 |
Online Publication Date | Dec 4, 2014 |
Publication Date | May 1, 2015 |
Deposit Date | Jul 7, 2015 |
Publicly Available Date | Jun 1, 2016 |
Journal | Radiography |
Print ISSN | 1078-8174 |
Publisher | Elsevier |
Peer Reviewed | Peer Reviewed |
Volume | 21 |
Issue | 2 |
Pages | 188 - 196 |
DOI | https://doi.org/10.1016/j.radi.2014.11.003 |
Keywords | Computed radiography |
Publisher URL | http://dx.doi.org/10.1016/j.radi.2014.11.003 |
Related Public URLs | http://www.sciencedirect.com/science/journal/10788174 |
Files
USIR_submission.pdf
(1.3 Mb)
PDF
Downloadable Citations
About USIR
Administrator e-mail: library-research@salford.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search