This article reports preliminary results of pilot studies of a new implantable two channel drop foot stimulator. The stimulator consists of an externally worn transmitter inductively coupled to an implanted receiver unit located in the lower leg, lateral and distal to the knee. The receiver is connected to electrodes located under the epineurium of the deep and the superficial peroneal nerves. Stimulation is triggered by detection of heel lift and terminated at heel strike in a manner similar to surface mounted systems. The location of the electrodes allows for a degree of selectivity over the resultant moment about the ankle joint that is not possible with surface stimulation of the common peroneal nerve. The two subjects used the stimulator on a regular basis and showed increases in walking speed of between 10% and 44% when compared to their baseline measurements. Isometric tests have demonstrated that the stimulator allows selective and repeatable stimulation of ankle joint muscles.
Keywords: Drop foot; Implant; Functional electrical stimulation; Stroke patients
Citation
Kenney, L., Bultstra, G., Buschman, R., Taylor, P., Mann, G., Hermens, H., …Hobby, J. (2002). An implantable two channel drop foot stimulator: initial clinical results. Artificial Organs, 26(3), 267-270. https://doi.org/10.1046/j.1525-1594.2002.06949.x
Journal Article Type | Article |
---|---|
Publication Date | Mar 1, 2002 |
Deposit Date | Aug 9, 2007 |
Journal | Artificial Organs |
Print ISSN | 0160-564X |
Publisher | Wiley |
Peer Reviewed | Peer Reviewed |
Volume | 26 |
Issue | 3 |
Pages | 267-270 |
DOI | https://doi.org/10.1046/j.1525-1594.2002.06949.x |
Publisher URL | http://dx.doi.org/10.1046/j.1525-1594.2002.06949.x |
You might also like
Why does my prosthetic hand not always do what it is told?
(2022)
Journal Article
Downloadable Citations
About USIR
Administrator e-mail: library-research@salford.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search