Skip to main content

Research Repository

Advanced Search

A phantom-based method to assess X-ray table mattress interface pressures

Alresheedi, N; Walton, AL; Tootell, AK; Webb, J; Hogg, P

A phantom-based method to assess X-ray table mattress interface pressures Thumbnail


Authors

N Alresheedi

AL Walton

AK Tootell

J Webb



Abstract

Background

Pressure redistribution performance of X-ray table mattresses can influence the development of pressure ulcers in at risk populations. Interface pressure analysis, with human participants, is a common method to assess mattresses. This approach has limitations that relate to the lack of standardisation between and within humans.

Aim

To develop and validate an anthropomorphic phantom-based method to assess X-ray table mattress interface pressures as an index of mattress performance.

Methods

A phantom simulating an adult head, pelvis and heels, was 3D printed from X-ray Computed Tomography image data and attached to a metal frame 175cm in length. Dry sand was added to the phantom head, pelvis and heels to represent a range of human weights. Pressure distribution was assessed using XSensor. Phantom validation was achieved by comparing phantom mattress interface pressure characteristics, for 5 human equivalent weights, against 27 sets of human mattress interface pressure data.

Results

Using the correlation coefficient R, phantom and human pressure data showed good correlation for the five phantom weights (R values: head=0.993, pelvis=0.997 and heels= 0.996).

Conclusion

A novel method to test X-ray mattresses for interface pressure was developed and validated. The method could have utility in the testing of X-ray mattresses that are in routine use and for new mattress development. Phantom interface pressure data could be provided by manufacturers to help inform procurement decisions when matching mattress characteristics to medical imaging demands and the underlying patient populations.

Citation

Alresheedi, N., Walton, A., Tootell, A., Webb, J., & Hogg, P. (2020). A phantom-based method to assess X-ray table mattress interface pressures. Journal of Medical Imaging and Radiation Sciences, 51(3), 417-424. https://doi.org/10.1016/j.jmir.2020.04.007

Journal Article Type Article
Acceptance Date Apr 22, 2020
Online Publication Date Jun 3, 2020
Publication Date Sep 1, 2020
Deposit Date Apr 22, 2020
Publicly Available Date Jun 3, 2021
Journal Journal of Medical Imaging and Radiation Sciences
Print ISSN 1939-8654
Publisher Elsevier
Volume 51
Issue 3
Pages 417-424
DOI https://doi.org/10.1016/j.jmir.2020.04.007
Publisher URL https://doi.org/10.1016/j.jmir.2020.04.007
Related Public URLs http://www.journals.elsevier.com/journal-of-medical-imaging-and-radiation-sciences/

Files






You might also like



Downloadable Citations