A novel method for comparing radiation dose and image quality, between and within different X-ray units in a series of hospitals

Al-Murshedi, S; Hogg, P; Lanca, L; England, A

A novel method for comparing radiation dose and image quality, between and within different X-ray units in a series of hospitals Thumbnail


Authors

S Al-Murshedi

L Lanca

A England



Abstract

Objectives: To develop a novel method for comparing radiation dose and image quality (IQ) to evaluate adult chest X-ray (CXR) imaging among several hospitals.
Methods: CDRAD 2.0 phantom was used to acquire images in eight hospitals (17 digital X-ray units) using local adult CXR protocols. IQ was represented by image quality figure inverse (IQFinv), measured using CDRAD analyser software. Signal to noise ratio (SNR), contrast to noise ratio (CNR) and conspicuity index (CI) were calculated as additional measures of IQ. Incident air kerma (IAK) was calculated using a solid-state dosimeter for each acquisition. Figure of merit (FOM) was calculated to provide a single estimation of IQ and radiation dose.
Results: IQ, radiation dose and FOM varied considerably between hospitals and X-ray units. For IQFinv, the mean (range) between and within the hospitals were 1.42 (0.83-2.18) and 1.87 (1.52-2.18), respectively. For IAK, the mean (range) between and within the hospitals were 93.56 (17.26 to 239.15) µGy and 180.85 (122.58-239.15) µGy, respectively. For FOM, the mean (range) between and within hospitals were 0.05 (0.01 to 0.14) and 0.03 (0.02-0.05), respectively.
Conclusions: The suggested method for comparing IQ and dose using FOM concept along with the new proposed FOM, is a valid, reliable and effective approach for monitoring and comparing IQ and dose between and within hospitals. It is also can be beneficial for the optimisation of X-ray units in clinical practice. Further optimisation for the hospitals /X-ray units with low FOM are required to minimise radiation dose without degrading IQ.

Citation

Al-Murshedi, S., Hogg, P., Lanca, L., & England, A. (2018). A novel method for comparing radiation dose and image quality, between and within different X-ray units in a series of hospitals. Journal of Radiological Protection, 38(4), https://doi.org/10.1088/1361-6498/aae3fa

Journal Article Type Article
Acceptance Date Sep 25, 2018
Online Publication Date Sep 25, 2018
Publication Date Oct 18, 2018
Deposit Date Oct 12, 2018
Publicly Available Date Sep 25, 2019
Journal Journal of Radiological Protection
Print ISSN 0952-4746
Electronic ISSN 1361-6498
Publisher IOP Publishing
Volume 38
Issue 4
DOI https://doi.org/10.1088/1361-6498/aae3fa
Publisher URL https://doi.org/10.1088/1361-6498/aae3fa
Related Public URLs http://iopscience.iop.org/journal/0952-4746

Files






You might also like



Downloadable Citations