Construction and validation of a low cost paediatric pelvis phantom

Ali, M; Hogg, P; Johansen, S; England, A

Construction and validation of a low cost paediatric pelvis phantom Thumbnail


Authors

M Ali

S Johansen

A England



Abstract

PURPOSE: Imaging phantoms can be cost prohibitive, therefore a need exists to produce low cost alternatives which are fit for purpose. This paper describes the development and validation of a low cost paediatric pelvis phantom based on the anatomy of a 5-year-old child.

METHODS: Tissue equivalent materials representing paediatric bone (Plaster of Paris; PoP) and soft tissue (Poly methyl methacrylate; PMMA) were used. PMMA was machined to match the bony anatomy identified from a CT scan of a 5-year-old child and cavities were created for infusing the PoP. Phantom validation comprised physical and visual measures. Physical included CT density comparison between a CT scan of a 5-year old child and the phantom and Signal to Noise Ratio (SNR) comparative analysis of anteroposterior phantom X-ray images against a commercial anthropomorphic phantom. Visual analysis using a psychometric image quality scale (face validity).

RESULTS: CT density, the percentage difference between cortical bone, soft tissue and their equivalent tissue substitutes were -4.7 to -4.1% and -23.4%, respectively. For SNR, (mAs response) there was a strong positive correlation between the two phantoms (r>0.95 for all kVps). For kVp response, there was a strong positive correlation between 1 and 8 mAs (r=0.85), this then decreased as mAs increased (r=-0.21 at 20 mAs). Psychometric scale results produced a Cronbach’s Alpha of almost 0.8.

CONCLUSIONS: Physical and visual measures suggest our low-cost phantom has suitable anatomical characteristics for X-ray imaging. Our phantom could have utility in dose and image quality optimisation studies.
Keywords: Pelvis phantom, low-cost, dose optimisation, validation, development.

Citation

Ali, M., Hogg, P., Johansen, S., & England, A. (2018). Construction and validation of a low cost paediatric pelvis phantom. European Journal of Radiology, 108, 84-91. https://doi.org/10.1016/j.ejrad.2018.09.015

Journal Article Type Article
Acceptance Date Sep 12, 2018
Online Publication Date Sep 14, 2018
Publication Date Nov 1, 2018
Deposit Date Sep 19, 2018
Publicly Available Date Sep 14, 2019
Journal European Journal of Radiology
Print ISSN 0720-048X
Publisher Elsevier
Volume 108
Pages 84-91
DOI https://doi.org/10.1016/j.ejrad.2018.09.015
Publisher URL https://doi.org/10.1016/j.ejrad.2018.09.015
Related Public URLs https://www.journals.elsevier.com/european-journal-of-radiology

Files

Construction and validation of a low cost paediatric pelvis phantom (to peter).pdf (768 Kb)
PDF





You might also like



Downloadable Citations