K Alzyoud
Impact of body part thickness on AP pelvis radiographic image quality and effective dose
Alzyoud, K; Hogg, P; England, A
Abstract
Introduction: Within medical imaging variations in patient size can generate challenges, especially when selecting appropriate acquisition parameters. This experiment sought to evaluate the impact of increasing body part thickness on image quality (IQ) and effective dose (E) and identify optimum exposure parameters.
Methods: An anthropomorphic pelvis phantom was imaged with additional layers (1 to 15 cm) of animal fat as a proxy for increasing body thickness. Acquisitions used the automatic exposure control (AEC), 100 cm source to image distance (SID) and a range of tube potentials (70 to 110 kVp). IQ was evaluated physically and perceptually. E was estimated using PCXMC software.
Results: For all tube potentials, signal to noise ratio (SNR) and contrast to noise ratio (CNR) deceased as body part thickness increased. 70 kVp produced the highest SNR (46.6 to 22.6); CNR (42.8 to 17.6). Visual grading showed that the highest IQ scores were achieved using 70 and 75 kVp. As thickness increases, E increased exponentially (r=0.96; p<0.001). Correlations were found between visual and physical IQ (SNR r= 0.97, p<0.001; CNR r=0.98, p<0.001).
Conclusion: To achieve an optimal IQ across the range of thicknesses, lower kVp settings were most effective. This is at variance with professional practice as there is a tendency for radiographers to increase kVp as thickness increases. Dose reductions were experienced at higher kVp settings and are a valid method for optimisation when imaging larger patients.
Citation
Alzyoud, K., Hogg, P., & England, A. (2019). Impact of body part thickness on AP pelvis radiographic image quality and effective dose. Radiography, 25(1), e11-e17. https://doi.org/10.1016/j.radi.2018.09.001
Journal Article Type | Article |
---|---|
Acceptance Date | Sep 6, 2018 |
Online Publication Date | Oct 3, 2018 |
Publication Date | Feb 1, 2019 |
Deposit Date | Sep 7, 2018 |
Publicly Available Date | Oct 3, 2019 |
Journal | Radiography |
Print ISSN | 1078-8174 |
Publisher | Elsevier |
Volume | 25 |
Issue | 1 |
Pages | e11-e17 |
DOI | https://doi.org/10.1016/j.radi.2018.09.001 |
Publisher URL | https://doi.org/10.1016/j.radi.2018.09.001 |
Related Public URLs | https://www.radiographyonline.com/ |
Files
Fat%20thickness%20article%20v_11%2027%208%20%2018%20ae%20ACCEPTED[1].pdf
(1.6 Mb)
PDF
You might also like
Advanced practice among diagnostic radiographers - An international survey
(2024)
Journal Article
Research activity among diagnostic and therapeutic radiographers: An international survey
(2024)
Journal Article
Digital Mammography: A Holistic Approach
(2022)
Book
Downloadable Citations
About USIR
Administrator e-mail: library-research@salford.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search