Impact of body part thickness on AP pelvis radiographic image quality and effective dose

Alzyoud, K; Hogg, P; England, A

Impact of body part thickness on AP pelvis radiographic image quality and effective dose Thumbnail


Authors

K Alzyoud

A England



Abstract

Introduction: Within medical imaging variations in patient size can generate challenges, especially when selecting appropriate acquisition parameters. This experiment sought to evaluate the impact of increasing body part thickness on image quality (IQ) and effective dose (E) and identify optimum exposure parameters.
Methods: An anthropomorphic pelvis phantom was imaged with additional layers (1 to 15 cm) of animal fat as a proxy for increasing body thickness. Acquisitions used the automatic exposure control (AEC), 100 cm source to image distance (SID) and a range of tube potentials (70 to 110 kVp). IQ was evaluated physically and perceptually. E was estimated using PCXMC software.
Results: For all tube potentials, signal to noise ratio (SNR) and contrast to noise ratio (CNR) deceased as body part thickness increased. 70 kVp produced the highest SNR (46.6 to 22.6); CNR (42.8 to 17.6). Visual grading showed that the highest IQ scores were achieved using 70 and 75 kVp. As thickness increases, E increased exponentially (r=0.96; p<0.001). Correlations were found between visual and physical IQ (SNR r= 0.97, p<0.001; CNR r=0.98, p<0.001).
Conclusion: To achieve an optimal IQ across the range of thicknesses, lower kVp settings were most effective. This is at variance with professional practice as there is a tendency for radiographers to increase kVp as thickness increases. Dose reductions were experienced at higher kVp settings and are a valid method for optimisation when imaging larger patients.

Citation

Alzyoud, K., Hogg, P., & England, A. (2019). Impact of body part thickness on AP pelvis radiographic image quality and effective dose. Radiography, 25(1), e11-e17. https://doi.org/10.1016/j.radi.2018.09.001

Journal Article Type Article
Acceptance Date Sep 6, 2018
Online Publication Date Oct 3, 2018
Publication Date Feb 1, 2019
Deposit Date Sep 7, 2018
Publicly Available Date Oct 3, 2019
Journal Radiography
Print ISSN 1078-8174
Publisher Elsevier
Volume 25
Issue 1
Pages e11-e17
DOI https://doi.org/10.1016/j.radi.2018.09.001
Publisher URL https://doi.org/10.1016/j.radi.2018.09.001
Related Public URLs https://www.radiographyonline.com/

Files

Fat%20thickness%20article%20v_11%2027%208%20%2018%20ae%20ACCEPTED[1].pdf (1.6 Mb)
PDF






You might also like



Downloadable Citations