Skip to main content

Research Repository

Advanced Search

A review of the design and clinical evaluation of the ShefStim array-based functional electrical stimulation system

Kenney, LPJ; Heller, B; Barker, A; Reeves, M; Healey, J; Good, T; Cooper, G; Sha, N; Prenton, S; Liu, A; Howard, D

A review of the design and clinical evaluation of the ShefStim array-based functional electrical stimulation system Thumbnail


Authors

B Heller

A Barker

M Reeves

J Healey

T Good

G Cooper

N Sha



Abstract

Functional electrical stimulation has been shown to be a safe and effective means of correcting foot drop of central neurological origin. Current surface-based devices typically consist of a single channel stim- ulator, a sensor for determining gait phase and a cuff, within which is housed the anode and cathode. The cuff-mounted electrode design reduces the likelihood of large errors in electrode placement, but the user is still fully responsible for selecting the correct stimulation level each time the system is donned. Researchers have investigated different approaches to automating aspects of setup and/or use, including recent promising work based on iterative learning techniques. This paper reports on the design and clini- cal evaluation of an electrode array-based FES system for the correction of drop foot, ShefStim. The paper reviews the design process from proof of concept lab-based study, through modelling of the array geome- try and interface layer to array search algorithm development. Finally, the paper summarises two clinical studies involving patients with drop foot. The results suggest that the ShefStim system with automated setup produces results which are comparable with clinician setup of conventional systems. Further, the final study demonstrated that patients can use the system without clinical supervision. When used un- supervised, setup time was 14 min (9 min for automated search plus 5 min for donning the equipment), although this figure could be reduced significantly with relatively minor changes to the de

Citation

Kenney, L., Heller, B., Barker, A., Reeves, M., Healey, J., Good, T., …Howard, D. (2016). A review of the design and clinical evaluation of the ShefStim array-based functional electrical stimulation system. Medical Engineering and Physics, 38(11), 1159 -1165. https://doi.org/10.1016/j.medengphy.2016.08.005

Journal Article Type Article
Acceptance Date Aug 13, 2016
Online Publication Date Sep 14, 2016
Publication Date Nov 1, 2016
Deposit Date Oct 31, 2016
Publicly Available Date Sep 14, 2017
Journal Medical Engineering & Physics
Print ISSN 1350-4533
Electronic ISSN 1873-4030
Publisher Elsevier
Volume 38
Issue 11
Pages 1159 -1165
DOI https://doi.org/10.1016/j.medengphy.2016.08.005
Publisher URL http://dx.doi.org/10.1016/j.medengphy.2016.08.005
Related Public URLs http://www.journals.elsevier.com/medical-engineering-and-physics/
Additional Information Funders : National Institute for Health Research (NIHR);Sheffield Hospitals Charitable Trust;UK Overseas Research Studentship