Background

This study reports the incidence and sequelae of migration of the Nellix (Endologix Inc, Irvine, Calif) endoprosthesis after endovascular aneurysm sealing.

Methods

A review was performed of the follow-up imaging of all endovascular aneurysm sealing patients in a university hospital endovascular program who had a minimum follow-up of 1 year. The first postoperative and latest follow-up computed tomography scans were used to measure the distances between the proximal and distal borders of the stent grafts relative to reference vessels using a previously validated technique. Device migration was based on previously established criteria and defined as any stent graft movement of ≥4 mm related to a predefined reference vessel. Device movement in a caudal direction was given a positive value, and movement in a cranial direction was denoted by a negative value.

Results

Eighteen patients (35 stent grafts) were eligible for inclusion in this retrospective review. The mean preoperative abdominal aortic aneurysm diameter was 57 mm (standard deviation [SD], 5; range, 50-67 mm) and aortic neck length was 30 mm (SD, 16; range, 6-62 mm). Proximal migration, according to study definitions, was identified in six stent grafts (17%), all in a caudal direction. At 1 year the mean proximal migration distance was +6.6 mm (SD, 1.6; range, +4.7-+9.2 mm). Migration occurred in a single stent graft in four patients and bilaterally in one. No distal migration occurred.

Conclusions

Proximal migration of the Nellix endoprosthesis does occur and was without any sequelae in our series. Further investigations into the long-term positional stability of the Nellix device, together with a more thorough understanding of the etiology and consequences of migration, are required.





Author conflict of interest: F.T. and R.F.K. have received professional fees and educational grants from Endologix.

The editors and reviewers of this article have no relevant financial relationships to disclose per the JVS policy that requires reviewers to decline review of any manuscript for which they may have a conflict of interest.

Citation

England, A., Torella, F., Fisher, R., & McWilliams, R. (2016). Migration of the Nellix Endoprosthesis. Journal of Vascular Surgery, 64(2), 306-312. https://doi.org/10.1016/j.jvs.2016.02.053

Journal Article Type Article
Acceptance Date Feb 23, 2016
Online Publication Date Apr 9, 2016
Publication Date Aug 1, 2016
Deposit Date Apr 26, 2016
Publicly Available Date Apr 9, 2017
Journal Journal of Vascular Surgery
Print ISSN 0741-5214
Electronic ISSN 1097-6809
Publisher Elsevier
Volume 64
Issue 2
Pages 306-312
DOI https://doi.org/10.1016/j.jvs.2016.02.053
Publisher URL http://dx.doi.org/10.1016/j.jvs.2016.02.053
Related Public URLs http://www.journals.elsevier.com/journal-of-vascular-surgery/

Files





You might also like



Downloadable Citations