Abstract
Background
This study reports the incidence and sequelae of migration of the Nellix (Endologix Inc, Irvine, Calif) endoprosthesis after endovascular aneurysm sealing.
Methods
A review was performed of the follow-up imaging of all endovascular aneurysm sealing patients in a university hospital endovascular program who had a minimum follow-up of 1 year. The first postoperative and latest follow-up computed tomography scans were used to measure the distances between the proximal and distal borders of the stent grafts relative to reference vessels using a previously validated technique. Device migration was based on previously established criteria and defined as any stent graft movement of ≥4 mm related to a predefined reference vessel. Device movement in a caudal direction was given a positive value, and movement in a cranial direction was denoted by a negative value.
Results
Eighteen patients (35 stent grafts) were eligible for inclusion in this retrospective review. The mean preoperative abdominal aortic aneurysm diameter was 57 mm (standard deviation [SD], 5; range, 50-67 mm) and aortic neck length was 30 mm (SD, 16; range, 6-62 mm). Proximal migration, according to study definitions, was identified in six stent grafts (17%), all in a caudal direction. At 1 year the mean proximal migration distance was +6.6 mm (SD, 1.6; range, +4.7-+9.2 mm). Migration occurred in a single stent graft in four patients and bilaterally in one. No distal migration occurred.
Conclusions
Proximal migration of the Nellix endoprosthesis does occur and was without any sequelae in our series. Further investigations into the long-term positional stability of the Nellix device, together with a more thorough understanding of the etiology and consequences of migration, are required.
Author conflict of interest: F.T. and R.F.K. have received professional fees and educational grants from Endologix.
The editors and reviewers of this article have no relevant financial relationships to disclose per the JVS policy that requires reviewers to decline review of any manuscript for which they may have a conflict of interest.
Citation
England, A., Torella, F., Fisher, R., & McWilliams, R. (2016). Migration of the Nellix Endoprosthesis. Journal of Vascular Surgery, 64(2), 306-312. https://doi.org/10.1016/j.jvs.2016.02.053
Journal Article Type | Article |
---|---|
Acceptance Date | Feb 23, 2016 |
Online Publication Date | Apr 9, 2016 |
Publication Date | Aug 1, 2016 |
Deposit Date | Apr 26, 2016 |
Publicly Available Date | Apr 9, 2017 |
Journal | Journal of Vascular Surgery |
Print ISSN | 0741-5214 |
Electronic ISSN | 1097-6809 |
Publisher | Elsevier |
Volume | 64 |
Issue | 2 |
Pages | 306-312 |
DOI | https://doi.org/10.1016/j.jvs.2016.02.053 |
Publisher URL | http://dx.doi.org/10.1016/j.jvs.2016.02.053 |
Related Public URLs | http://www.journals.elsevier.com/journal-of-vascular-surgery/ |
Files
manuscript to endologix.pdf
(507 Kb)
PDF
Licence
http://creativecommons.org/licenses/by-nc-nd/4.0/
Publisher Licence URL
http://creativecommons.org/licenses/by-nc-nd/4.0/
You might also like
A narrative review on pressure ulcer (PU) studies relevant to medical imaging
(2020)
Journal Article
Low-dose chest CT and the impact on nodule visibility
(2020)
Journal Article
Inclusion of evidence and research in European radiography curricula
(2020)
Journal Article
Downloadable Citations
About USIR
Administrator e-mail: library-research@salford.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search