Dose optimisation in paediatric radiography - using regression models to investigate the relative impact of acquisition factors on image quality and radiation dose

Ali, MA; Hogg, P; England, A

Authors

MA Ali

A England



Abstract

Objective: To investigate the optimum pelvis X-ray acquisition factors for a 10-year-old child. Secondly, to evaluate the impact of each acquisition factor on image quality (IQ) and radiation dose.

Method: Images were acquired using a pelvis phantom and a range of acquisition parameters; e.g. tube potential, additional filtration and source-to-image distance (SID). Automatic exposure control (AEC) was used with two orientations (head towards/away from two outer chambers) and three different chamber selections. Visual IQ was evaluated using relative and absolute-VGA methods. Radiation doses were measured by placing a dosimeter on the anterior surface of the phantom. Regression analysis was used to determine optimum parameters.

Results: The optimised technique (178.8 µGy), with diagnostic IQ, was with 89kVp, 130 cm SID and with 1 mm Al + 0.1 mm Cu filtration. This technique was with the head towards the two outer AEC chambers. Regression analysis showed that SID had the lowest impact on IQ (β = 0.002 95% CI −0.001 to 0.005) and dose (β = −0.96 95% CI −0.40 to −1.53). The impact of filtration on dose (β = −76.24 95% CI −86.76 to −85.72) was higher than tube potential (β = −13.44 95% CI −14.34 to −12.53). The following impact ratios were higher on IQ than radiation dose: filtration/kVp; 11.28 times, filtration/SID; 7.01 times and kVp/SID; 0.62 times.

Conclusion: Optimised parameters were identified as 89 kVp, 130 cm SID and with 1 mm Al + 0.1 mm Cu additional filtration. Regression analysis demonstrated that filtration and tube potential had the greatest effect on radiation dose and IQ, respectively.

Citation

Ali, M., Hogg, P., & England, A. (2019). Dose optimisation in paediatric radiography - using regression models to investigate the relative impact of acquisition factors on image quality and radiation dose. Physica Medica, 68, 61-68. https://doi.org/10.1016/j.ejmp.2019.10.034

Journal Article Type Article
Acceptance Date Oct 21, 2019
Online Publication Date Nov 18, 2019
Publication Date Nov 18, 2019
Deposit Date Jul 22, 2019
Publicly Available Date Nov 21, 2019
Journal Physica Medica : European Journal of Medical Physics
Print ISSN 1120-1797
Electronic ISSN 1724-191X
Publisher Elsevier
Volume 68
Pages 61-68
DOI https://doi.org/10.1016/j.ejmp.2019.10.034
Publisher URL https://doi.org/10.1016/j.ejmp.2019.10.034
Related Public URLs https://www.sciencedirect.com/journal/physica-medica

Files






You might also like



Downloadable Citations