The readout thickness versus the measured thickness for a range of screen film mammography (SFM) and full-field digital mammography (FFDM) units
Hauge, IHR; Hogg, P; Szczepura, K; Connolly, P; McGill, G; Mercer, C
Authors
Prof Peter Hogg P.Hogg@salford.ac.uk
Ms Katy Szczepura K.Szczepura@salford.ac.uk
Associate Professor/Reader
P Connolly
G McGill
C Mercer
Abstract
PURPOSE: To establish a simple method to determine breast readout accuracy on mammography units. METHODS: A thickness measuring device (TMD) was used in conjunction with a breast phantom. This phantom had compression characteristics similar to human female breast tissue. The phantom was compressed, and the thickness was measured using TMD and mammography unit readout. Measurements were performed on a range of screen film mammography (SFM) and full-field digital mammography (FFDM) units (8 units in total; 6 different models/manufacturers) for two different sized paddles and two different compression forces (60 N and 100 N).
RESULTS: The difference between machine readout and TMD for the breast area, when applying 100 N compression force, for non-flexible paddles was largest for GE Senographe DMR+ (24 cm x 30 cm paddle: +14.3%). For flexible paddles the largest difference occurred for Hologic Lorad Selenia (18 cm x 24 cm paddle: +26.0%).
CONCLUSIONS: None of the units assessed were found to have perfect correlation between measured and readout thickness. TMD measures and thickness readouts were different for the duplicate units from two different models/manufacturers.
Citation
Hauge, I., Hogg, P., Szczepura, K., Connolly, P., McGill, G., & Mercer, C. (2011). The readout thickness versus the measured thickness for a range of screen film mammography (SFM) and full-field digital mammography (FFDM) units. Medical Physics, 39(1), 263-271. https://doi.org/10.1118/1.3663579
Journal Article Type | Article |
---|---|
Acceptance Date | Nov 2, 2011 |
Online Publication Date | Dec 20, 2011 |
Publication Date | Dec 20, 2011 |
Deposit Date | Nov 3, 2011 |
Publicly Available Date | Nov 3, 2011 |
Journal | Medical Physics |
Print ISSN | 0094-2405 |
Publisher | American Association of Physicists in Medicine |
Peer Reviewed | Peer Reviewed |
Volume | 39 |
Issue | 1 |
Pages | 263-271 |
DOI | https://doi.org/10.1118/1.3663579 |
Publisher URL | http://dx.doi.org/10.1118/1.3663579 |
Related Public URLs | http://online.medphys.org/ |
Files
26-10-2011_revision_2.docx
(2.2 Mb)
Document
You might also like
The student perspective of a research-informed teaching activity using simulation
(2024)
Presentation / Conference
Skin tears in mammography: A narrative review
(2023)
Journal Article
Downloadable Citations
About USIR
Administrator e-mail: library-research@salford.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search